The Nitrocellulose membrane is the most pivotal part of a rapid test kit as this is where the antigen-antibody binding occurs. Manufactured under strictly controlled conditions, Nupore membranes are cast on transparent polyester backing and offer highly consistent results, reliable inter-lot performance, and sharp control/test lines with minimal non-specific background signal.
| NC Membrane Type | LFNC 70 | LFNC 15µm | LFNC 12µm | LFNC 10µm | LFNC 140 | LFNC 180 |
|---|---|---|---|---|---|---|
| Backing | 100µm polyster | 100µm polyster | 100µm polyster | 100µm polyster | 100µm polyster | 100µm polyster |
| Membrane Thickness | 95 - 135 µm | 95 - 135 µm | 95 - 135 µm | 95 - 135 µm | 95 - 135 µm | 95 - 135 µm |
| Wicking Rate (sec/4cm) | 64 - 90 | 70 - 100 | 80 - 110 | 95 - 135 | 120 - 160 | 160 - 200 |
All Nupore NC membranes are white and have a smooth and flat surface with no macroscopic defects or foreign matter
| NC Membrane Type | LFNC 70 | LFNC 15µm | LFNC 10µm | LFNC 140 |
|---|---|---|---|---|
| Sensitivity | Low | Medium | High | Very High |
| Test Time | Low | Medium | High | Very High |
Nupore NC membranes are recommended to be stored in original foil pouch at an ambient temperature (around 20°C) for longer shelf life. As the membrane is inflammable, it should be kept away from direct heat and sunlight.